Can placing lead-rubber inferolateral to the light beam diaphragm limit ionising radiation to multiple radiosensitive organs?

Hayre, Christopher, Bungay, H, Jeffery, C, Cobb, Christopher and Atutornu, Jerome (2017) Can placing lead-rubber inferolateral to the light beam diaphragm limit ionising radiation to multiple radiosensitive organs? Radiography, 24 (1). pp. 15-21. ISSN 1078-8174

[thumbnail of Can placing lead-rubber inferolateral to the light beam diaphragm limit ionising radiation to multiple radiosensitive o.docx] Text
Can placing lead-rubber inferolateral to the light beam diaphragm limit ionising radiation to multiple radiosensitive o.docx - Accepted Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

Download (3MB)

Abstract

Introduction

This article investigates a practical method of reducing the impact of scattered radiation during a lateral radiographic projection of the elbow. The light beam diaphragm (LBD) is generally accepted to limit ionising radiation using horizontal and longitudinal lead shutters, yet this article evidences further dose limitation by placing lead-rubber inferolateral to the LBD device.

Methods

Using an anthropomorphic phantom and arm construction scattered radiation was recorded at multiple radiosensitive organs. A 15 cc ionisation chamber (model 10100 AT TRIAD) was placed on each radiosensitive organ (eye, thyroid, breast, testes, spleen and ovaries) measuring exposure rate (μGy/s). Dose readings were recorded before and after the placement of lead-rubber inferolateral to the LBD. A paired two sample t-test was undertaken affirming how likely dose limitation was attributable to chance (p < 0.05).
Results

Descriptive and inferential statistics demonstrate dose reduction to radiosensitive organs (right eye 53%, right breast 53%, left eye 39%, thyroid 13%, left ovary 9%, testes 6%, left breast 3% and spleen 2%) upon placement of the lead-rubber inferolateral to the LBD. The paired two sample t-test demonstrated statistically significant dose limitation (t = 2.04, df = 7, p = 0.04) thus significant for radiographic practice.
Conclusion

Placement of lead-rubber inferolateral to the LBD limits dose to multiple radiosensitive organs. Right (53%) and left (39%) eye lens, right breast (53%), thyroid (13%), left ovary (9%), testes (6%), left breast (3%) and spleen (2%) statistically demonstrate dose limiting opportunities to patients.

Item Type: Article
Uncontrolled Keywords: X-ray experiment, Dose limitation, Light beam diaphragm, ALARP, General radiography
Subjects: R Medicine > R Medicine (General)
Divisions: Faculty of Health & Science > Department of Health Studies
Depositing User: David Upson-Dale
Date Deposited: 22 Nov 2017 08:45
Last Modified: 17 Jul 2019 08:33
URI: https://oars.uos.ac.uk/id/eprint/268

Actions (login required)

View Item
View Item